Manual

page 29 of 35
Fault Possible cause Remedy
No aspiration.
Filter clogged?
Replace lter.
Leak at the bottle head?
Check seals, coupling
and blind cap. Screw in
or replace if necessary.
No vacuum supply con
nected?
Connect vacuum supply.
No vacuum while vacu
um supply connected?
Check vacuum supply.
Open lock valve in the
vacuum line if applica�
ble.
Pressure control
device switches fre
quently.
Leak in the system?
Check tubing, seals,
coupling, and blind cap.
Screw in or replace if nec
essary.
Filter clogged?
Replace lter.
Vacuum controller con�
taminated inside or
defective?
Contact local distributor.
Troubleshooting