The use of the disinfectant Sekusept
®
Plus
(Manufacturer: Ecolab GmbH & Co OHG, Düs
seldorf, Germany) in extensive in-house test
ing did not cause any damage to the collect
ing bottle. Hence the disinfectant Sekusept
®
Plus - even when used in the collecting bottle
during suction - is preferred to chlorine bleach.
Comply with use and safety instructions of the
manufacturer!
+ Even if using disinfecting solutions after termination of the suction it
is absolutely necessary to assure the compatibility of disinfectant and
the parts to be disinfected.
+ For information about the compatibility with the materials of the pump
ing unit, ask the manufacturer of the disinfectant.
+ The wetted materials of the system are listed in the section ”Technical
data”.
Sekusept
®
Plus