Instruction Manual

page 22 of 29
Fault Possible cause Remedy
No aspiration at
the VHC
pro
. Pipette
drips.
Adapter or aspiration
tube defective?
Replace adapter with
aspiration hose.
Aspiration hose kinked?
Correct hose guide.
Aspiration hose clogged
or agglutinated?
Remove clogged or ag-
glutinated hose section
or replace hose.
Pipette ts loosely or ts
bad?
Connect pipette secure-
ly or choose an appro-
priate adapter.
Leak in the system?
Check the system for
leaks.
No vacuum?
Check the vacuum sys-
tem.
Pipette does not
held properly in the
adapter.
Adapter wet or greasy?
Clean or replace adapter.
Pipette with improper
diameter selected?
Choose appropriate pi-
pette or change adapter.
Seal at the nose cone
defectice?
Replace nose cone.
Liquid is aspirated
slowly or not.
Pipette or nose cone
clogged or damaged?
Clean or replace pi-
pettes or nose cones.
Troubleshooting