Specifications

Page: 16
INTERNAL FUNCTIONS
(F8) Tare Mode Selection
a) Use the procedure described above to reach Function F8.
b) Press MODE/RANGE to enter, and display will show
“X-tArE” (X being the previous selected mode)
n_tArE Tare function can be used when weight is not
stable
S_tArE Tare function can be used only when weight
is stable
c) Press MODE/RANGE to select the desired setting.
d) Press TARE to confirm, then press ON/ZERO to return to
normal weighing mode.
(F9) Sound Alarm for HI-OK-LO
a) Use the procedure described above to reach Function F9.
b) Press MODE/RANGE and the display will show XXbEEP
(XX being the setting from the last selection)
nobEEP disable sound alarm
gobEEP sound alarm when OK value is reached
HLbEEP sound alarm when HI/LO value is reached
HbEEP sound alarm when HI value is reached
LbEEP sound alarm when LO value is reached
c) Press MODE/RANGE to select the desired setting.
Page: 17
INTERNAL FUNCTIONS
d) Press TARE to confirm, then press ON/ZERO to return to
normal weighing mode.
(F10) Relay Output Mode Selection
a) Use the procedure described above to reach Function F10.
b) Press MODE/RANGE to enter function.
c) Use MODE/RANGE to choose either Act L or Act H.
d) Press TARE to confirm, then press ON/ZERO to return to
normal weighing mode.
Normal Action
Act L
Relay output status
is open
Relay output status
is closed
Act H
Relay output status
is closed
Relay output status
is open