Service Procedures
Return Procedure
A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service prior to
returning any product.
Replacement Parts and Accessories
To order accessories or replacement parts for the GelDoc-It
e
Imager, contact UVP’s offices.
Part Description Part Number
Fuses:
Fuse, 3.2A (for Darkroom) 56-0002-05 (Qty. 2 Required)
Fuse, 2A (for 115V Transilluminators) 56-0002-01 (Qty. 2 Required)
Fuse, 2A (for 230V Transilluminators) 56-0002-03 (Qty. 2 Required)
Emission Filters:
Filter, Ethidium Bromide, 50mm Square 38-0220-01
Filter, SYBR Green, 50mm Square 38-0219-01
Filter, SYBR Gold, 50mm Square 38-0221-01
Transillumination Accessories:
White Light Converter Plate, 21x26cm 38-0191-01
Visi-Blue Converter Plate, 21x26cm 38-0200-01
Gel Accessories:
Gel-Cutter 85-0002-01
Gel-Ruler 85-0003-01
Gel-Scooper 85-0006-01
Gel-Tray, small 85-0007-01
Gel-Sentry DNA Preparation Plate 97-0076-01
Fluorescent Standard Step Tablet 33-0014-02
Protective Equipment:
Spectacles, UV Blocking (UVC-303) 98-0002-01
Goggles, UV Blocking (UVC-503) 98-0002-02
Faceshield, UV Blocking (UVC-803) 98-0002-04
Thermal Printer and Paper:
Thermal Printer 89-0069-06
High Gloss, High Contrast Thermal Paper (5 Rolls per Box) 89-0174-01
Matte Finish, High Contrast Thermal Paper (4 Rolls per Box) 89-0038-01
Troubleshooting
No Power to the Darkroom or Transilluminator
1. Recheck the main power cord connection to the GelDoc-It
e
darkroom as well as the jumper cable
between the darkroom and transilluminator.
2. Check the fuses located at the back of the unit next to the power port. A flat-head screwdriver will
be required.
Turn the fuseholder cap counterclockwise and the fuse holder will pop out. Inspect the thin wire
within the glass fuse to see if there is a break in the wire. If so, replace the fuse(s). If fuses are
blowing repeatedly, contact UVP Technical Support for additional troubleshooting.