Operating instructions

With Benchtop Transilluminator and Drawer
When using the drawer and Benchtop Transillumina-
tor with the DigiDoc-It System, a +2 diopter is required
for accurate camera focus. Your system may have a
diopter attached to the camera lens if you are using
the system with a larger High Performance Transillu-
minator and no drawer. The following instructions are
for replacing the diopter.
Diopter Installation
Check the camera lens for an existing diopter and UV
filter. If installed, remove both. The diopter shipped
with the drawer has two diopters assembled at the
UVP factory. Screw the diopter assembly onto the
camera lens. Replace the UV filter.
Drawer Installation
Place the Benchtop Transilluminator into the drawer.
Connect the power cord to the back of the transillu-
minator and push the plug through the access hole in
the back of the drawer. Connect the plug to the power
source. Access to the transilluminator switch is via
the access port on the front of the drawer.
Place the DigiDoc-it hood enclosure over the drawer.
Camera Set-Up
You may remove the camera from the hood to snap
images of subjects other than gels or samples. You
can easily replace the camera in the hood as follows:
Insert the Camera into the Hood
1. Open the camera cap on top of the darkroom hood.
2. Position the camera on top of the darkroom hood
so that the cameras lens fits into the hole on the
darkroom hood.
3. Insert the knob screw through the cameras mount-
ing bracket and turn the knob screw counter-clock-
wise to tighten it into the cameras tripod mounting
threads.
4. Plug the cameras power cable into the input power
receptacle on the camera.
5. Make sure the cables pass through the notched
hole in the camera cap provided for them.
4
DigiDoc-It Systems
DigiDoc-It with drawer
Hood Cap