User Manual
GelDoc-It
TS2
and ChemiDoc-It
TS2
Imagers 27
Service Procedures
Return Procedure
A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.
Replacement Parts and Accessories
To order accessories or replacement parts for the GelDoc-It
TS2
and ChemiDoc-It
TS2
Imagers, contact UVP’s
offices.
Part Description Part Number
Epi UV Modules:
UV Module, 254/365nm, 4 watt (115V) 95-0021-12 (Qty. 2 Recommended)
UV Module, 254/365nm, 4 watt (230V) 95-0021-10 (Qty. 2 Recommended)
Fuses:
Fuse, 3.15A (for Darkroom) 56-0022-04 (Qty. 2 Required)
Fuse, 2A (for Transilluminators) 56-0002-01 (Qty. 2 Required)
Emission Filters:
Filter, Ethidium Bromide, 50mm Square 38-0220-01
Filter, SYBR Green, 50mm Square 38-0219-01
Filter, SYBR Gold, 50mm Square 38-0221-01
Transillumination Accessories:
LED White Light Plate 95-0476-01
White Light Converter Plate, 21x26cm 38-0191-01
Visi-Blue Converter Plate, 21x26cm 38-0200-01
Gel Accessories:
Gel-Cutter 85-0002-01
Gel-Ruler 85-0003-01
Gel-Scooper 85-0006-01
Gel-Tray, small 85-0007-01
Gel-Sentry DNA Preparation Plate 97-0076-01
Fluorescent Standard Step Tablet 33-0014-02
Protective Equipment:
Spectacles, UV Blocking (UVC-303) 98-0002-01
Goggles, UV Blocking (UVC-503) 98-0002-02
Faceshield, UV Blocking (UVC-803) 98-0002-04
Troubleshooting
No Power to the Darkroom or Transilluminator
1. Recheck the main power cord connection to the GelDoc-
It
TS2
/ChemiDoc-It
TS2
darkroom as well as the power cables
between the darkroom and transilluminator, LED White Light
Plate or optional epi UV modules.
2. Check the fuses located at the back of the unit next to the
power port. A small flat-head screwdriver or similar tool is
required. Push the bottom tab of the fuseholder up until the
bottom of the fuseholder pops out. Then, push the top tab
down until the top pops out. The entire fuseholder can now be
removed.