User Guide
2 3
Figure 2.
Figure 4.
Figure 5.
Figure 3.
Pulsed-Xenon unit angled towards bed. Ideal placement for dosimeter is circled.
Correct – Chair has been moved no longer obstructing light. Dosimeter on side of
ling cabinet has been removed as well.
Three dosimeter cards prepared for testing on bed.
Incorrect - Objects obstructing light and improper placement of dosimeter card.
2) Prepare selected surfaces in step one for testing. Remove anything that may obstruct
the light pathway to the target surface. Avoid shadowing and indirect exposure to
dosimeter card surface.
3) Remove the 3M protective strip from the top of the back of the cards to expose
adhesive. Place cards adhesive side down on target surfaces selected in step one.
a. At minimum place one card per target surface.
b. To understand irradiance over time we recommend three cards per target
surface.
Dosimeter on side of filing
cabinet will receive indirect light
and provide low readings.
Chair will obstruct light from
pulsed-Xenon unit, casting shadow
over cards and lower readings.