User's Manual

A canned CO testing agent must be used for the CO functional gas
test.
If a tamper alarm occurs:
Make sure that you are using a compatible control panel (see
Specificationsbelow).
1. Wait at least 10 minutes after installation to test the CO
alarm.
Make sure that the control panel is in sensor test mode
during sensor testing.
2. Make sure the green Power LED is flashing for normal
operation.
Alarm/open condition does not restore:
3. Set the control panel to sensor test mode.
Make sure that the CO alarm condition has cleared at the CO
alarm.
4. Press and hold the Test/Hush button until the unit beeps
three times (approximately 10 seconds), and then release the
button. The unit will enter the functional gas test mode. The
Power LED will blink once per second while in functional test
mode.
Make sure that you are using a compatible control panel (see
Specificationsbelow).
Specifications
5. Apply UL approved CO test agent. When CO is detected, the
unit will activate a CO alarm. The unit will send RF test
packets to the control panel when the CO alarm is activated.
Compatible panels Simon XT (firmware 1.4 or greater)
Power Three AA batteries
Battery type 1.5 VDC alkaline
Required batteries Duracell MN 1500, Duracell MX1500, Energizer
E91
Sensor Electrochemical
Sensor life 7 years
Frequency 319.5 MHz
Supervisory interval 64 minutes
Audible alarm Temporal 4
Alarm response times 70 PPM = 60-240 min.
150 PPM = 10-50 min.
400 PPM = 4-15 min.
Dimensions 4.68 x 2.75 x 1.85 in. (119 x 70 x 47 mm)
Storage temperature -4 to 140°F (-20 to 60°C)
Operating environment
Temperature
Relative humidity
40 to 100°F (4.4 to 37.8°C)
10 to 95% noncondensing
6. The control panel will beep and display the number of RF
packets received.
7. At the control panel, exit sensor test mode.
8. Exit functional gas test mode:
Press and release the Test/Hush button; or
A 2 minute timeout will automatically cause the CO to return
to normal operating mode.
Troubleshooting
This information is provided to help you diagnose and solve
various problems that may arise while configuring or using the
wireless CO alarm.
Unit does not power up properly or reports low battery:
Make sure the batteries are fully seated within the battery
compartment and the polarity is correct.
Make sure that all three batteries are installed.
Check the battery voltage (1.5 VDC nominal per battery).
Regulatory information
Control panel does not respond:
Use the 60-401 RF Sniffer to make sure the CO alarm is
sending messages for activation.
ETL rating ANSI/UL 2034
FCC compliance This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference,
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
FCC ID: B4Z-929A-CO
Move or rotate the CO alarm position.
Make sure the CO alarm is properly enrolled into the control
panel.
Make sure that you are using a compatible control panel (see
Specificationsbelow).
Tamper condition does not restore:
Make sure the CO alarm body is properly installed on the
mounting plate.
Make sure there are no trouble indications at the CO alarm.
Make sure that you are using a compatible control panel (see
Specificationsbelow).
4 / 8 P/N 466-2393 • REV 00.15 • ISS 07DEC10
Draft
IC: 1175C-929ACO
This Class B digital apparatus complies with
Canadian ICES-003.
Cet appareil numérique de la classe B est
conforme à la norme NMB-003 du Canada.
Changes or modifications not expressly approved by the party responsible
for compliance could void the user's authority to operate the equipment.
In accordance with FCC requirements of human exposure to radio
frequency fields, the radiating element shall be installed such that a
minimum separation distance of 20 cm is maintained from the general
population.