Operating instructions
LIFEPAK
®
500
automated external defibrillator
O
PERATOR’S CHECKLIST
Unit Serial No.:
Location:
Instruction
Recommended
Corrective Action
Date
Initials
1 Examine the AED case, connector,
and battery well for:
Foreign substances Clean the device.
Damage or cracks Contact authorized service personnel.
2 Examine the battery pins for bending
or discoloration.
Discard and replace battery.
3 Check expiration date on batteries
and therapy electrodes.
Replace if expired.
4 Examine the accessory cables for
cracked, damaged, broken, or bent
connectors or wires.
Replace damaged or broken parts.
5 With the battery installed, press
On/Off to turn on the AED and look
for:
Self-test messages If absent, contact authorized service
personnel.
Momentary illumination of each LED
and all LCD segments
If absent, contact authorized service
personnel to repair or replace parts.
BATTERY LOW or REPLACE BATTERY
SELF-TEST XX.XX
message
Replace the battery immediately.
Service indicator or
CALL SERVICE
message
Contact authorized service personnel.
I
n
s
e
r
t
a
✔
i
n
t
h
e
b
o
x
a
f
t
e
r
c
o
m
p
l
e
t
i
n
g
e
a
c
h
i
n
s
t
r
u
c
t
i
o
n
.
This is a suggested checklist for inspecting and
checking this device on a daily basis and after each
use. You may also consult JAMA, August 22/29,
1990, Vol. 264, No. 8, Table 3 for the Defibrillator
Working Group's automated defibrillator checklist.
This form may be reproduced.