Operating instructions
6-6 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Troubleshooting
Table 6-6 LIFEPAK 500 AED Voice Prompts
REPLACE BATTERY
The battery voltage is very low.
SELF-TEST XX.XX
The self-test is being performed and software version xx.xx is installed.
SEND COMPLETE
The AED successfully transferred data.
SENDING
The AED is transferring the setup to another AED.
SENDING
XX% COMPLETE
The AED is transferring data by modem or to a printer. The transfer is xx%
complete.
SETUP MODE
nnnnnnnnnnnnnnnnnnn
The AED is in the setup mode. The nnnnnnnnnnnnnnnnnnn is the Device
Configuration code.
SHOCK ADVISED
The AED has analyzed the patient ECG rhythm and detected a shockable
ECG rhythm.
STAND CLEAR
The AED prompt to move everyone away from the patient.
START CPR
The AED prompt that follows the IF NO PULSE message.
STOP MOTION
See MOTION DETECTED.
TEST MODE
The AED has entered the test mode.
TEST OK
The external test load test has been successfully completed.
TO PRINT PUSH " The AED is connected to a printer and ready to print a report.
TO SEND PUSH " The AED is connected to a modem and ready to transfer data.
TRANSFER SETUP TO
SEND PUSH "
Setup mode message for the Transfer Setup feature.
TRY AGAIN
The AED is ready for you to retry transferring data by modem.
WILL RE-DIAL IN
XX SECONDS
While attempting to transfer data by modem, the AED detected that the
destination phone number was busy. The AED will try again in xx seconds.
Voice Prompt Description
ANALYZING NOW, STAND
CLEAR
The AED is analyzing the patient ECG rhythm.
ASYSTOLE
The AED has analyzed the patient ECG and detected persistent asystole.
CHECK FOR PULSE
Check the patient for a pulse.
CHECK FOR SIGNS OF
CIRCULATION
AED prompt after each standard three-shock sequence or NO SHOCK
ADVISED
message when PULSE PROMPT 2 is selected in Setup.
CONNECT ELECTRODES
The AED detects that the electrodes are disconnected.
IF NO PULSE, START CPR
If patient pulse is not present, start CPR.
IF NO PULSE, PUSH ANALYZE
If patient pulse is not present, press ANALYZE.
IF NO SIGNS OF CIRCULATION
AED prompt that follows the CHECK PATIENT prompt when PULSE
PROMPT 2 is selected in Setup.
MOTION DETECTED, STOP
MOTION
The AED detects motion during ECG analysis.
NO SHOCK ADVISED
The AED has analyzed the patient ECG and detected a non-shockable
ECG rhythm.
PUSH ANALYZE
Press ANALYZE to begin ECG analysis.
REPLACE BATTERY
The battery voltage is low and must be replaced immediately.
SHOCK ADVISED
The AED has analyzed the patient ECG and detected a shockable ECG
rhythm.
Screen Message Description