Operating instructions

LIFEPAK 500 Automated External Defibrillator Operating Instructions 6-5
©1996–2003 Medtronic Physio-Control Corp.
Troubleshooting
6 Troubleshooting
Table 6-5 LIFEPAK 500 AED Screen Messages
Receiving AED failed to receive
transmission.
Turn the receiving AED off and on with
the Setup Transfer Cable connected.
Screen Message Description
ANALYZING NOW
The AED is analyzing the patient ECG rhythm.
ASYSTOLE
The AED has analyzed the patient’s ECG and detected persistent asystole.
ASYSTOLE DETECTOR
Setup mode message for asystole time option.
AUDIO RECORDING
Setup mode message for the audio recording option.
AUTO ANALYZE
Setup mode message for auto analyze options.
BATTERY OK
The battery voltage is ok.
BUSY
While attempting to transfer data by modem, the AED detected that the
destination phone number was busy.
CALL SERVICE
The AED detected a fault requiring immediate service during self-tests.
CANNOT SEND
The AED could not transfer the setup, print a report, or transfer data through
a modem.
CHARGE REMOVED
The SHOCK button has been disarmed.
CHECK FOR PULSE
AED prompt after each standard three-shock sequence or NO SHOCK
ADVISED
message when PULSE PROMPT 1 is selected in Setup.
CHECK FOR SIGNS OF
CIRCULATION
AED prompt after each standard three-shock sequence or NO SHOCK
ADVISED message when PULSE PROMPT 2 is selected in Setup.
CONNECT ELECTRODES
The AED has detected that the electrodes are disconnected.
CPR TIME
XX SEC
Setup mode message for the CPR timer option.
CPSS DURING CPR
Setup mode message for CPSS during CPR option.
DEVICE ID
XXXXXXXXX
Setup mode message for device ID option.
ENERGY SEQUENCE
#2-XXX
Setup mode message for energy sequence option.
IF NO PULSE
AED prompt that follows the CHECK FOR PULSE message.
IF NO SIGNS OF CIRCULATION
AED prompt that follows the CHECK FOR SIGNS OF CIRCULATION prompt
when
PULSE PROMPT 2 is selected in Setup.
LOW BATTERY
The battery voltage is low.
MODEM INIT STRING
XXXXXXXXXXXXXXXXXXXXXX
Setup mode message for the modem initialization string option.
MODEM PHONE NUMBER
XXXXXXXXXXXXXXXXXXXXXXX
Setup mode message for the modem phone number option.
MODEM SELECTION
# XX
Setup mode message. You may select the configuration for one of nine
Hayes AT-compatible modems.
MOTION DETECTED
The AED detects motion during ECG analysis, thereby inhibiting analysis.
MOTION DETECTION
Setup mode message for motion detection option.
NO SHOCK ADVISED
The AED has analyzed the patient ECG and detected a nonshockable ECG
rhythm.
PUSH ANALYZE
Press ANALYZE to begin ECG analysis.
PUSH TO SHOCK
The AED is fully charged and ready to provide therapy. This is the AED
prompt to press
SHOCK to discharge.
Observation Possible Cause Corrective Action