6-4 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Troubleshooting
Table 6-3 Troubleshooting During Printing
Table 6-4 Troubleshooting During Setup Transfer
4 LOW BATTERY
message or
indicators
appear on Readiness
Display and key panel.
Low battery. If using AED, continue to use and
replace battery at earliest convenience.
Approximately 20% of battery energy
remains.
If AED not is use, replace the battery
immediately.
5 REPLACE
BATTERY
voice prompt
or indicator on
key panel flashes.
Very low battery. Replace battery immediately.
Observation Possible Cause Corrective Action
1 Printer does not print. No printer power. Make sure the printer cord is
connected.
Make sure the printer switch is on.
No printer paper. Check the printer paper.
Printer cable not connected. Check the printer cable connections.
Wrong type of printer. Check the printer to make sure that it is
EPSON ESC/P-compatible.
2 Printed report does not line
up properly on paper.
Wrong paper size selected. Make sure correct paper size is
selected (8 1/2 x 11 or A4) in
SETUP
OPTIONS
menu.
3 CONNECT ELECTRODES
message:
The " button was not held down
when the AED was turned on.
Hold down the " button while turning
on the AED.
4 LOW BATTERY
message or
indicators
appear on Readiness
Display and key panel.
Low battery. If using AED, continue to use and
replace battery at earliest convenience.
Approximately 20% of battery energy
remains.
If AED not in use, replace the battery
immediately.
5 REPLACE
BATTERY
voice prompt
or indicator
on key panel flashes.
Very low battery. Replace battery immediately.
Observation Possible Cause Corrective Action
Original AED displays
CANNOT SEND message.
Setup Transfer Cable is not
properly connected.
Check the connections between the
Setup Transfer Cable, the original AED,
and the receiving AED.
Wrong cable is connected. Connect the Setup Transfer Cable to
the original AED and the receiving
AED.
Receiving AED is not on. Make sure the receiving AED is on.
Receiving AED was turned on
with electrodes connected or
while AED was connected to
computer or modem.
Turn the receiving AED off and on with
the Setup Transfer Cable connected.
Observation Possible Cause Corrective Action