LIFEPAK 500 Automated External Defibrillator Operating Instructions 6-3
©1996–2003 Medtronic Physio-Control Corp.
Troubleshooting
6 Troubleshooting
Table 6-2 Troubleshooting During Modem Data Transfer
SHOCK button not pressed
within 15 seconds.
•Press SHOCK within 15 seconds after
the
PUSH TO SHOCK message appears.
9 Displayed time is incorrect. Time is incorrectly set in the
AED.
Change the AED time setting.
10 Date printed on report is
incorrect.
Date is incorrectly set in the
AED.
Change the AED date setting.
11 Displayed messages are
faint or flicker.
Low battery power.
Out of Temperature Range.
Replace the battery immediately.
12 Voice prompts sound faint
or distorted.
Low battery power. Replace the battery immediately.
13 AED operates but LCD is
blank.
Operating temperature is too
low or too high.
Operate the AED between 0
° and 50°C
(32
° to 122°F).
LCD not operating properly. Contact authorized service personnel.
14 AED turns off or will not
turn on.
Depleted battery. Replace the battery immediately.
Disconnected battery. Install battery.
Observation Possible Cause Corrective Action
1 BUSY and WILL RE-DIAL IN
XX SECONDS
messages.
Destination number is busy, the
AED is preparing to retry.
Wait for the AED to retry the data
transfer.
AED will retry up to three times.
2 TRY AGAIN, TO SEND PUSH
or
CANNOT SEND
messages.
Transmission failed. AED will retry up to three times.
Wrong phone number. Check the destination phone number
and
MODEM PHONE NUMBER setup
option.
Cable is not properly connected. Check connections.
Modem is not connected to an
analog telephone line.
Confirm that the telephone line is
analog (not digital).
Incorrect modem selected in
Setup menu.
Check modem selected in
SETUP
OPTIONS
menu.
Custom Modem Init String is
incorrect.
•Check
MODEM INIT STRING.
Dial string for destination site is
incorrect.
Check the AED
MODEM PHONE
NUMBER setup option.
Computer power at destination
is not on.
Make sure the computer power is on.
Computer application program
is not ready.
Make sure the program is ready to
receive data.
Connection failed or is busy.
AED has tried to send data three
times.
Resend the data.
3 CONNECT ELECTRODES
message.
AED was turned on before
modem.
Turn off the AED for one minute. Then,
turn on the modem
before
the AED
power and resend the data.
Observation Possible Cause Corrective Action