Operating instructions

5-16 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Maintenance
Shock MIL-STD-810E, Method 516.4, Procedure 1 (40 g, 6–9 ms pulse, 1/2 sine
each axis).
Vibration Monophasic version: MIL-STD-810E, Method 514.4, Category 10 Biphasic
version: MIL-STD-810E, Method 514.4, Helicopter - Category 6
(3.75 Grms) and Ground Mobile - Category 8 (3.15 Grms). RTCA D0 160C,
Table 8-2 Fixed Wing - Turbojet Engine Classification C’ (Fuselage). Test
level per Figure 8-5 C’. One hour in each of three axes.
Aircraft
Tested to RTCA/DO-160C, “Environmental Conditions and Test
Procedures for Airborne Equipment.” (Details available on request.)
*Note: See page 5-7 for information on caring for batteries.
GENERAL
Rechargeable SLA battery pak
Ty pe
Sealed lead-acid, 8V, 2.5 amp hours.
Capacity
Typical: 59 full discharges or 3 hours of “on time” with a new, fully charged
battery.
Minimum: 43 full discharges with a new, fully charged battery.
Battery Charge Time
10 ±1 hours. Battery charging limited to 15° to 35°C (59° to 95°F).
Recommended
Replacement Interval
2 years or 200 battery charge/discharge cycles, whichever comes first
using recommended battery maintenance procedures.
Weight
0.9 kg (1.9 lb).
Nonrechargeable battery pak
Lithium sulfur dioxide
(LiSO
2
) battery
Ty pe
Sealed lithium, 12V, 7.5 amp-hours.
Certification
FAA: TSO – C97 or CAA: BS2G239.
Capacity
Typical: 312 full discharges or 14 hours of “on time” with a new battery.
Minimum: 230 full discharges with a new battery.
Shelf Life
5 years (4 years for aircraft use).
Weight
0.5 kg (1.2 lb)
Lithium manganese
dioxide (LiMnO
2
) battery
Ty pe
Sealed lithium, 12V, 10 amp-hours.
Capacity
Typical: 416 full discharges or 18 hours of “on time” with a new battery.
Minimum: 306 full discharges with a new battery.
Shelf Life
5 years
Weight
0.5 kg (1.2 lb)
Physical Characteristics
Height
10.2 cm (4.0 in)
Width
26.7 cm (10.5 in)