LIFEPAK 500 Automated External Defibrillator Operating Instructions 5-15
©1996–2003 Medtronic Physio-Control Corp.
Maintenance
5 Maintenance
Service Indicator Service icon.
Displayed Messages Messages prompt user through complete operating sequence.
Audible Tones Coded tones assist user through device operation and alert operator of
display messages.
Voice Prompts Prompt user through complete operation sequence.
EVENT DOCUMENTATION
Type Internal digital memory.
Memory Capacity 20 minutes audio recording (optional).
ECG and event log of operator/device actions:
At least 20 minutes if unit is configured with audio recording and audio
recording setup option is
ON.
At least 80 minutes if configured with audio recording and audio recording
setup option is
OFF.
At least 60 minutes if not configured with audio recording.
Report Types CODE SUMMARY report, Event Log report, Test Log report.
Capacity 300 Event Log events.
30 Test Log device tests (assuming no fault codes).
Communications Options:
– Direct connection to personal computer
– Modem connection to personal computer using Hayes AT-Compatible
modem
– Print direct with EPSON
®
ESC/P protocol for printers with 9-pin
printheads.
Data Review
LIFENET
™
system compatible. Options:
– DATA TRANSFER
™
500 information management program
– QUIK-VIEW
™
500 data review program
– CODE-STAT
™
Suite data management system, v 2.0 or above.
ENVIRONMENTAL
Operating Temperature 0° to 50°C* (32° to 122°F).
Storage Temperature
-30° to 65°C* (-22° to 149°F) without battery and electrodes.
-30° to 65°C* (-22° to 149°F) with battery and electrodes, maximum total
exposure time limited to one week.
Atmospheric Pressure 760 to 429 mmHg (0 to 15,000 ft above sea level).
Relative Humidity 10 to 95% (non-condensing).
Water Resistance IEC 60529/EN 60529 IPX4 “Splash-proof” with electrodes or connector
cover installed.