5-14 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Maintenance
SPECIFICATIONS
Table 5-4 lists the specifications for the LIFEPAK 500 AED (monophasic and biphasic).
Table 5-5 lists the specifications for the LIFEPAK 500 AED Battery Charger.
Table 5-4 LIFEPAK 500 AED Specifications
1
AED
Input ECG via QUIK-COMBO or FAST-PATCH disposable electrodes. Standard
placement: anterior-lateral. Alternate placement: anterior-posterior.
Electrical Protection Input protected against high voltage defibrillator pulses per IEC 60601/
EN 60601.
Safety Classification Internally powered equipment IEC 60601-1/EN 60601-1, 5.1.
Waveform Monophasic pulse (Edmark) per AAMI DF2-1989, 3.2.1.5.1
Biphasic truncated exponential, with voltage and duration compensation for
patient impedance.
2
Output Energy Accuracy ±15% into 50 ohms (monophasic)
±10% into 50 ohms (biphasic)
±15% into 25 to 100 ohms (biphasic)
Output Energy Sequence Monophasic: 200, 200, 360 joules (360 joules thereafter) or 200, 300, 360
joules (360 joules thereafter).
Biphasic: Three levels, user configurable from 200 to 360 joules, delivered
(Level 1, Level 2, Level 3, Level 3...).
Charge Time With a new, nonrechargeable battery pak, or a new, fully charged
rechargeable battery pak:
200 joules in less than 9 seconds
360 joules in less than 15 seconds.
Controls
ON/OFF
Turns device power on or off.
ANALYZE
Starts ECG analysis. (Optional).
SHOCK
Delivers defibrillation energy. Active only when Shock Advisory System
advises defibrillation.
Clock Set
Two switches
and are provided to set the clock.
Display Two-line, 20-character per line dot matrix liquid crystal display.
Readiness Display Indicates
OK when self-test completed successfully.
Low Battery Indicator Low battery icon:
At least 11 discharges remaining with nonrechargeable battery pak
At least 6 discharges remaining with rechargeable battery pak.
1
All specifications at 20°C (68°F) unless otherwise stated. All performance specifications assume the device has been
stored (two hours minimum) at the operating temperature prior to operation.
2
Specifications apply from 25 to 200 ohms. Voltage compensation is limited to the voltage that would result in delivery of
360 joules into 50 ohms.