Operating instructions

5-2 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Maintenance
MAINTENANCE AND TESTING SCHEDULING
The LIFEPAK 500 AED performs an automatic self test every 24 hours. If the automatic self test detects
a low battery condition or a condition that requires service, the monophasic AED activates an alarm;
AEDs with a readiness display will change the indicators on the display and will not activate an alarm
unless
AUDIO ALERT is configured ON. It is important to place the AED where the alarm is likely to be
heard and to periodically inspect the AED. Periodically check display indicators on AEDs with a
readiness indicator (refer to the following inspection subsections).
The AED also performs a self-test every time you turn on the AED. These self-tests do not eliminate the
need for regular maintenance. You should do the following on a regular basis and after each time the
AED is used:
Inspect the AED as described in Table 5-1.
Clean the AED as described in Table 5-2.
Check to make sure that all necessary supplies and accessories (such as properly-maintained
batteries and therapy electrodes) are readily accessible.
When establishing your local operator maintenance schedule, consider how often the AED is used and
how familiar the operators are with AED operation. For example:
If the AED is used on a weekly basis, daily inspections may be appropriate.
If the AED is used on a monthly basis, weekly inspections may be appropriate.
If the AED is used very infrequently, such as once a year, monthly inspections may be appropriate.
INSPECTION
Routinely inspect all devices, accessories, and cables by following the instructions in Table 5-1.
Table 5-1 LIFEPAK 500 AED Inspection
Instruction Inspect for Recommended Corrective Action
Examine the AED case,
connector, battery well,
battery pins, and
accessories.
Foreign substances. Clean the device as described in Table 5-2.
Damage or cracks. Contact authorized service personnel to
troubleshoot and repair parts.
Battery pins bent or discolored. Contact authorized service personnel to
replace or repair parts.
Expired batteries or defibrillation
electrodes.
Replace.
AEDs with Readiness
Display:
Observe Readiness Display
OK
None needed.
Battery indicator displayed Replace battery immediately.
Service indicator displayed Contact authorized service personnel to
replace or repair parts.