4-6 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Data Management
7 After a few seconds, you will see the message:
TO SEND PUSH "
•Press " to send the Current Patient.
•Press
! to send the Previous Patient.
• Press both
" and ! to send the Current and Previous Patients.
8 If the incident ID option is
ON and an Incident ID has
not
already been entered for the Current or
Previous Patient, you will see the message:
ENTER CURRENT [ or PREVIOUS] ID?
YES
•Press ANALYZE (or the blank “menu” button) to answer YES; or
•Press
! to change to NO. Then press ANALYZE (or the blank “menu” button) and continue with step
10.
9 If you answered
YES, you will see the message:
INCIDENT ID
XXXXXXXXX
•Press ! to scroll through and select from the alphanumeric characters available.
•Press
" to advance to the next field.
• Repeat this process until the Incident ID is entered.
•Press
ANALYZE (or the blank “menu” button) to accept the Incident ID.
Note: The last Incident ID entered will always be displayed.
10 Verify Incident ID entered. You will see the message:
XXXXXXXXX
OK TO SEND? YES
•Press ANALYZE (or blank “menu” button) to accept and send the Incident ID.
•Press
! to change to NO.
•Press
ANALYZE (or blank “menu” button) to return to Incident ID screen.
• Follow step 9 beginning with bulleted items to change the Incident ID.
11 After
ANALYZE (or the blank “menu” button) is pressed, the AED transfers the patient data. While the
data is being transferred, the AED displays the following message to indicate progress:
SENDING
XX%COMPLETE
After the AED successfully completes the data transfer, it displays the SEND COMPLETE message.
12 After the AED displays the
SEND COMPLETE message, check that the low battery indicator is not
displayed.
13 Turn off the AED and prepare it for the next patient use.
Note: If you leave the LIFEPAK 500 AED unattended during data transfer, the AED automatically
turns off after 15 minutes of no activity (after data transfer completed).
If the AED turns off, check the data transfer status:
1 Leave the data cable connected to AED and modem.
2 Turn on AED and look for the
SEND COMPLETE message. If the CANNOT SEND message appears, refer
to Table 6-2 on page 6-3 for troubleshooting tips.
3 If the
SEND COMPLETE message appears:
• Check that the low battery indicator is not displayed.
• Disconnect the data transfer cable.
4 Turn off the AED and prepare it for the next patient use.
Troubleshooting During Data Transfer
If you cannot transfer data, refer to Table 6-2 on page 6-3 for troubleshooting tips.