Operating instructions

3-8 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Using the LIFEPAK 500 AED
After CPR Time
After CPR TIME, the AED displays one of the following message sequences, depending on the PULSE
PROMPT
selected in Setup:
You will hear the corresponding voice prompt. Within 10 seconds, the AED displays two alternating
messages if
AUTO ANALYZE is off or AUTO ANALYZE 1 is selected:
You will hear the corresponding voice prompt.
Note: If
AUTO ANALYZE 2 is selected or for LIFEPAK 500 AEDs that do not have an ANALYZE button,
analysis will begin automatically at the end of CPR Time. You will hear
STAND CLEAR, ANALYZING
NOW, STAND CLEAR. Stop CPR immediately and stay clear of patient during the analysis.
Shock Counter
The AED displays the shock counter in the upper-left corner of the LCD:
The shock counter indicates how many shocks have been delivered to the patient. Following the shock
counter, the energy for that shock number may be displayed (optional). The shock counter resets
whenever the AED is turned off for at least 60 seconds.
Motion Detection
If the AED is configured with MOTION DETECTION ON and the AED detects motion during the ECG
analysis, the AED alternately displays two messages:
CHECK FOR PULSE
10:52
or
CHECK FOR SIGNS
10:52
OF CIRCULATION
10:52
IF NO PULSE
10:52
PUSH ANALYZE
10:52
or
IF NO SIGNS
10:52
OF CIRCULATION
10:52
PUSH ANALYZE
09:29
#1
Shock counter
Energy of shock #1
200J
MOTION DETECTED
09:44
STOP MOTION
09:44
#2
#2