2-16 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Getting Ready
3 Turn on the receiving AED and wait for CONNECT ELECTRODES message to appear.
4 Push the
" button on the original AED to send the setup to the receiving AED.
During setup transfer, the original AED displays the
SENDING message. The receiving AED
displays a blank screen.
After the original AED successfully transfers the setup, it displays the
SEND COMPLETE message.
The receiving AED turns itself off, turns itself back on, and then displays the
CONNECT
ELECTRODES
message.
5 To transfer the setup from the original AED to additional AEDs:
• Turn off the receiving AED.
• Disconnect the Setup Transfer Cable from the receiving AED.
• Repeat Steps 2 through 4.
6 When finished, disconnect the Setup Transfer Cable, turn off both AEDs, and prepare them for
patient use.
Note: The original AED does not transfer the device ID to the receiving AED. To change the
device ID on a receiving AED, refer to the LIFEPAK 500 automated external defibrillator Setup
Instructions (PN 3012275).
CONNECTING ELECTRODES TO THE AED
You can connect the QUIK-COMBO electrodes with the REDI-PAK preconnect system to the AED
before patient care to save time. To connect the REDI-PAK-type QUIK-COMBO electrodes:
1 Inspect the electrode package and confirm that the expiration date has not passed.
2 Remove the clear plastic pouch to expose the QUIK-COMBO electrode connector.
3 Open the connector cover on the AED as shown in Figure 2-5.
4 Insert the electrode connector firmly into the cable connector on the AED as shown in Figure 2-5.
Figure 2-5 Connecting the QUIK-COMBO electrodes
5 Store the electrodes in the carrying case or the electrode storage tray.
6 Do not open the electrode package until immediately prior to patient use.
If you use QUIK-COMBO electrodes without the preconnect system, you should:
• Not open the electrode package until immediately prior to patient use.
• Inspect the electrode package and confirm that the expiration date has not passed.
• Store the electrode package in the carrying case or electrode storage tray.
• When ready for patient use, open the electrode package and connect the electrodes to the AED as
shown in Figure 2-5 above.
Note: If you are using FAST-PATCH electrodes, refer to Appendix C.
Connector Cover
Cable Connector
QUIK-COMBO