Operating instructions
LIFEPAK 500 Automated External Defibrillator Operating Instructions 2-15
©1996–2003 Medtronic Physio-Control Corp.
Getting Ready
2 Getting Ready
21 Press ANALYZE (or blank “menu” button) to advance to the AUDIO ALERT screen. (This screen appears
only on AEDs distributed in the English language):
• Press the
▲ button to change the setting (choices are ON/OFF).
22 Press
ANALYZE (or blank “menu” button) to advance to the TRANSFER SETUP screen:
• To transfer the setup to another AED, refer to Transferring Setup to Another LIFEPAK 500 AED,
which follows.
– or –
To turn off the AED, press
ON/OFF. The settings are saved.
TRANSFERRING SETUP TO ANOTHER LIFEPAK 500 AED
You can transfer the clock setting and all setup information except DEVICE ID from one LIFEPAK 500
AED to an identical AED using the Transfer Setup option. Identical AEDs are devices that have the
same button configuration, software version, and defibrillation waveform.
Note: Only LIFEPAK 500 AEDs with software version 4.2 or later can transfer and receive setup
data. Attempting to transfer setup data to devices with software version 4.0 may induce erroneous
faults in the receiving device.
To transfer the setup:
1 From within the
SETUP MODE, push ANALYZE (or blank “menu” button) to advance to the transfer setup
option. The AED displays the
TRANSFER SETUP screen:
2 Connect the equipment as shown in Figure 2-4:
• Connect the Setup Transfer Cable to the AED that has the setup you wish to transfer (original
AED).
• Connect the other end of the Setup Transfer Cable to the AED that you wish to receive the new
setup (receiving AED).
Note: Both AEDs must have the same button configuration and defibrillation waveform.
Figure 2-4 Setup transfer connections
AUDIO ALERT
OFF
TRANSFER SETUP
TO SEND PUSH
TRANSFER SETUP
TO SEND PUSH
Setup
Transfer Cable
LIFEPAK 500 AED
LIFEPAK 500 AED