Operating instructions
2-12 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Getting Ready
CHANGING SETUP OPTIONS
To change the setup options:
1 Make sure the LIFEPAK 500 AED power is off for at least 60 seconds and that nothing is connected
to the AED.
2 If the LIFEPAK 500 AED has an
ANALYZE button, hold down the ANALYZE, ▲, and buttons. Then,
press
ON/OFF. Do not release ANALYZE, ▲, and until the SETUP MODE message appears.
If the LIFEPAK 500 AED does not have an
ANALYZE button, hold down the blank “menu” button
(located between the
ON/OFF and SHOCK buttons), ▲, and buttons. Then press ON/OFF. Do not
release the blank “menu” button,
▲, and until the SETUP MODE message appears.
3 Notice that the AED displays the
SETUP MODE screen:
The
nnnnnnnnnnnnnnnnnnn is the configuration code. This code, which appears at the top of each
printed report, summarizes some of the setup and service settings.
Note: A letter B in the first space of the configuration code indicates that the device delivers a
biphasic defibrillation waveform.
4 Press
ANALYZE (or blank “menu” button) to advance to the DEVICE ID screen:
• Press the
▲ button to change the character (choices are 0-9, A-Z, #-., and space characters).
• Press the button to advance to the next space.
5 Press
ANALYZE (or blank “menu” button) to advance to the MODEM PHONE NUMBER screen:
• Press the
▲ button to change the character. The characters available are: ()*,- 0 through
9 PTW#$.
• Press the button to advance to the next character location (20 characters maximum).
6 Press
ANALYZE (or blank “menu” button) to advance to the MODEM SELECTION screen:
• Press the
▲ button to change the modem selection (select 0, 1, 2, 3, 4, or 5).
7 Press
ANALYZE (or blank “menu” button) to advance to the MODEM INIT STRING screen if the MODEM
SELECTION
is 0 (or advance to the ENERGY SEQUENCE screen if the MODEM SELECTION is 1, 2, 3, 4, or
5):
• Press the
▲ button to change the character (choices are all displayable characters).
• Press the button to advance to the next character location (75 characters maximum).
Call your Medtronic Physio-Control service representative for assistance with defining the correct
initialization string.
SETUP MODE
nnnnnnnnnnnnnnnnnnn
DEVICE ID
– – – – – – – – – – – – – – –
MODEM PHONE NUMBER
T9W18886279698
MODEM SELECTION
0
MODEM INIT STRING
– – – – – – – – – – – – – – –