LIFEPAK 500 Automated External Defibrillator Operating Instructions 2-5
©1996–2003 Medtronic Physio-Control Corp.
Getting Ready
2 Getting Ready
ABOUT BATTERIES
Use either of the following battery types to power the LIFEPAK 500 AED:
LIFEPAK 500 rechargeable sealed lead-acid (SLA) battery pak
LIFEPAK 500 nonrechargeable lithium sulfur dioxide (LiSO
2
) battery pak
LIFEPAK 500 nonrechargeable lithium manganese dioxide (LiMnO
2
) battery pak
To save battery life if the LIFEPAK 500 AED is accidentally turned on or left on, the AED has a battery
conservation feature. If the AED is not connected to a patient and no buttons are pressed for 15
minutes, the AED will automatically turn off.
With a battery installed, the LIFEPAK 500 AED automatically performs daily auto tests when the AED is
not in use. These auto tests, along with normal battery self-discharge, consume battery energy.
For information about maintaining or recharging the batteries, refer to page 5-7.
Battery Installation
To install a battery:
1 Insert the connector end of the battery into the battery compartment as shown in Figure 2-3.
2 Slide the battery all the way in until it latches securely.
Figure 2-3 Battery installation
Battery Removal
To remove the battery:
1 Turn off the AED.
2 Lift the latch release on the battery and slide it out.
Note: When a battery is removed from the AED, battery and service indicators appear on the
Readiness Display. After replacing the battery, turn on the device to reset the Readiness Display.
WARNING!
Inability to provide therapy.
The LIFEPAK 500 nonrechargeable lithium manganese dioxide battery pak does not fit in all LIFEPAK
500 AEDs. Use only with AEDs marked -003 inside the battery well.
Latch release
0
0