Operating instructions

LIFEPAK 500 Automated External Defibrillator Operating Instructions 1-3
©1996–2003 Medtronic Physio-Control Corp.
Safety Information
1 Safety Information
Note: Medtronic Physio-Control devices, electrodes, and cables are latex-free.
SYMBOLS
The symbols below may be found in this manual or on various configurations of the LIFEPAK 500 AED
and accessories:
WARNINGS!
Possible improper device performance.
Using other manufacturers’ cables, electrodes, or batteries may cause the device to perform improperly
and invalidates the safety agency certification. Use only the accessories specified in these Operating
Instructions.
Safety risk and possible equipment damage.
Monitors, defibrillators, and their accessories (including electrodes and cables) contain ferromagnetic
materials. As with all ferromagnetic equipment, these products must not be used in the presence of the
high magnetic field created by a Magnetic Resonance Imaging (MRI) device. The high magnetic field
created by an MRI device will attract the equipment with a force sufficient to cause death or serious
personal injury to persons between the equipment and the MRI device. This magnetic attraction may
also damage the equipment. Skin burns will also occur due to heating of electrically conductive
materials, such as patient leads and pulse oximeter sensors. Consult the MRI manufacturer for more
information.
Shock hazard.
Do not insert a hand, foot, or any object other than a battery into the battery well of this device.
CAUTION!
Possible equipment damage.
This device may be damaged by mechanical or physical abuse such as immersion in water or dropping
the device. If the device has been abused, remove it from use and contact a qualified service
technician.
Defibrillation protected, type BF patient connection
Attention, consult accompanying documents
Warning, high voltage
Indicator, steady display indicates battery is low, replace battery; flashing (key panel
only) indicates replace battery immediately
Indicator, steady display indicates device requires service; flashing (key panel only)
indicates service is required immediately
Indicator, appears on Readiness Display indicating self-test completed successfully
Buttons for setting the clock, transferring data, and setting options
Type BF patient connection
OK