Operating instructions

QUIK-COMBO
defibrillation cable
for LIFEPAK
®
500 automated external defibrillator
Instructions for Use (continued)
D-2 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Cleaning and Testing
To clean the QUIK-COMBO defibrillation cable,
wipe the surface with any one of the following:
Mild soap and water
Isopropyl alcohol
Peracetic (peroxide) acid solutions
Quaternary ammonium compounds
Contact local infection control resources for
specific questions regarding cleaning procedures
or cleaning agents available in your area.
Do not immerse or soak the defibrillation cable.
Do not use bleach or bleach dilution.
Do not steam or gas sterilize.
Inspect and test the defibrillation cable on a
routine basis. Inspection and testing will help
ensure that the equipment is in good operating
condition and is ready for use when needed. Use
the QUIK-COMBO Patient Simulator to test the
defibrillation cable.
During inspection or testing, if any discrepancy is
detected with the defibrillation cable, remove the
defibrillation cable from use and immediately
contact a qualified service representative.
Recycling Information
Recycle the device at the end of its useful life.
Preparation
The device should be clean and contaminant-free
prior to being recycled.
Recycling Assistance
The device should be recycled according to
national and local regulations. Contact your local
Medtronic Physio-Control representative for
assistance.
Recycling of Disposable Electrodes
After disposable electrodes are used, follow your
local clinical procedures for recycling.
Packaging
Packaging should be recycled according to
national and local regulations.
Ordering Information
Contact your local Medtronic Physio-Control
sales or service office to order parts. In the USA,
call the Medtronic Physio-Control PARTSLINE
at 1.800.442.1142.
QUIK-COMBO defibrillation cable kit for
LIFEPAK 500 AED (PN 3011215)
QUIK-COMBO Patient Simulator
(PN 803499-09)
QUIK-COMBO pacing/defibrillation/ECG
electrodes (2 ft lead wire) (PN 806086)
QUIK-COMBO LLW pacing/defibrillation/ECG
electrodes (3.5 ft lead wire) (PN 3008826)
QUIK-COMBO pacing/defibrillation/ECG
electrodes with REDI-PAK preconnect system
(PN 3008497)