Instructions for Use (continued)
C-2 LIFEPAK 500 Automated External Defibrillator Operating Instructions
Connecting to FAST-PATCH
defibrillation/ECG electrodes
Properly connect the defibrillation cable to the
electrodes to help ensure energy delivery (see
Figure 4).
• Attach the cable to the electrode post before
attaching electrodes to the patient.
• Support the electrode post under the electrode
when attaching the cable to the electrode.
• Firmly press the snap connector onto the
electrode post until a click is heard or felt.
• Confirm a secure connection of the cable to the
electrode before proceeding with therapy by
pulling up gently on the snap connector.
Figure 4 Connecting to FAST-PATCH defibrillation/ECG
electrodes
Note: If reattaching an electrode that is already
on the patient, lift the adhesive edge under the
electrode post slightly and place your finger
under the post. Connect the cable as described
above.
Disconnecting from disposable
electrodes
Disconnect the defibrillation cable from the
electrode by pulling the snap connector straight
up and off the post to avoid damage to the cable
or the post (see Figure 5).
Figure 5 Disconnecting from electrodes
Colors and Symbols
The defibrillation cable has colors and symbols
on the snap connectors consistent with industry
standards:
• AHA standards - red and white
• IEC standards - green and red
The snap connectors are labeled “+” (apex) and
“–” (sternum). Refer to the FAST-PATCH
electrode operating instructions for electrode
placement information.
Cleaning and Testing
To clean the FAST-PATCH defibrillation cable
and snap connectors, wipe the surface with any
one of the following:
• Mild soap and water
• Isopropyl alcohol
• Peracetic (peroxide) acid solutions
• Quaternary ammonium compounds
Contact local infection control resources for
specific questions regarding cleaning procedures
or cleaning agents available in your area.
• Do not immerse or soak the defibrillation cable
• Do not use bleach or bleach dilution
• Do not steam or gas sterilize
Inspect and test the defibrillation cable on a
routine basis. Inspection and testing will help
ensure that the equipment is in good operating
condition and is ready for use when needed. Use
the Physio-Control Patient Simulator to test the
defibrillation cable.
During inspection or testing, if any discrepancy is
detected with the defibrillation cable, remove the
defibrillation cable from service and immediately
contact a qualified service representative.
Ordering Information
Contact your local Physio-Control sales or
service office to order parts. In the USA, call the
Physio-Control PARTSLINE at 1.800.442.1142.
• FAST-PATCH defibrillation cable for LIFEPAK
500 AED (PN 3010493)
• FAST-PATCH disposable defibrillation/ECG
electrodes (PN 3006292)
• PHYSIO-CONTROL Patient Simulator
(PN 803499)