Specification Sheet

Advanced surface mount technology automatically compensates for ambient changes in the environment ensuring accuracy and consistency
in performance. Field-cleanable sensor is able to withstand 400 PSI of positive pressure. Unit operates 25 hours on 9V alkaline battery
with built in low battery indicator and auto-off feature. Units include a carrying care, hanging hook, AVT45 brass adapter and 9V battery.
ELECTRONIC VACUUM GAUGES
SENSOR TYPE
Self-heated thermistor bridge with integral temperature compensation from 0 to 50 °C
(32 to 122 °F)
OPERATING PRESSURE RANGE Atmospheric pressure to 1 micron. Maximum Working pressure: 400 PSIG
BURST PRESSURE 3000 PSIG
OPERATING RANGE Atmospheric to 1 micron
ACCURACY +/- 10% of Reading
OPERATING TEMPERATURE RANGE
Compensated (accuracy as stated): 0°C TO 50°C (32°F TO 122°F) Non-compensated
(add +/- 0.5% error for every °C outside compensated range): 0 to 50°C
(-22 to 158°F)
OPERATING & STORAGE HUMIDITY 0-95%, non-condensing
STORAGE TEMPERATURE - 40 to 185°F / - 40 to 85°C
POWER SOURCE One 9 volt alkaline battery (included)
BATTERY LIFE 25 hours of continuous use. Auto-off after 10 minutes
WEIGHT 5.5 oz. / 153 g. excluding the battery
DIMENSIONS
68 mm. diameter by 40 mm. deep by 124 mm. overall height (2.7" dia. by 1.6" deep by
5" overall height)
MECHANICAL CONNECTION Standard 1/4" female SAE refrigerant hose type with core depressor
BATTERY INDICATOR Battery symbol with 3 segment Power Level Bar Graph (VG200)
WORLD HEADQUARTERS
1010 East 31st Street, Hialeah, Florida 33013, USA
Tel: 305-687-4121, 1-800-277-3808, Fax: 305-687-3743
e-mail: info@cpsproducts.com www.cpsproducts.com
LTVG11
VG100A
Nine discrete high intensity LED’s
are preset to illuminate upon
achieving industry standards for
proper system evacuations.
VG200
Advanced measurement instrument
providing continual digital readout of
evacuation process on large 5 digit LCD.
Measurement units are user selectable
to read in Microns, Torr, in/hg or mBar
at the touch of a key.