PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 87
Anatel Operator Manual
Run standards button takes you to the run standards dialog box. See Run standards,
page 87.
Grab sample button takes you to the grab sample load bottles dialog box. See Grab sample,
page 104.
Excursion mode button takes you to the excursion mode dialog box. See Excursion mode,
page 108.
Unload bottles button takes you to the unload/replace bottles dialog box. See Figure 6-12,
page 95.
6.4 Run standards
The run standards dialog box allows you to select a standards test to run. See Figure 6-2.
Figure 6-2 Run standards dialog box
TOC calibration button takes you to the TOC calibration setup dialog box.
Conductivity calibration button takes you to the conductivity calibration load bottles dialog
box. See Conductivity calibration, page 96.
TOC validation button takes you to the TOC validation setup dialog box. See TOC
validation, page 100.
System suitability button takes you to the system suitability load bottles dialog box. See
System suitability test, page 102.
Return icon returns you to the bottle mode dialog box. See Bottle mode dialog box, page 86.
NOTE
Immediately following each test the standards bottles are backflushed and allowed to
equilibrate to ambient pressure. DO NOT ATTEMPT TO REMOVE THE BOTTLES UNTIL
THE BACKFLUSH AND PRESSURE EQUALIZATION IS COMPLETE.