User manual
Page 78 Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
TOC validation tab
Figure 5-26 Bottle test setup dialog box, TOC validation tab
Return to online mode checkbox enables the analyzer to return to the online mode
following the validation. You may enable or disable (default) this option. When enabled, the
analyzer returns to online mode following the validation. When disabled, the analyzer stays at
the bottle mode dialog box. All valves are closed, the pump is off, and the lamp is off.
Return icon returns you to the setup dialog box and saves any changes. Any changes are
noted in the audit trail.
System suitability tab
Figure 5-27 Bottle test setup dialog box, system suitability tab
Return to online mode checkbox enables the analyzer to return to the online mode
following the system suitability test. You may enable or disable (default) this option. When
enabled, the analyzer returns to online mode following the system suitability. When disabled,