User manual
PAT700 Total Organic Carbon Analyzer – Setup Page 75
Anatel Operator Manual
Filter icon opens the filter data dialog box. See Figure 5-23.
Return icon returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail.
Figure 5-22 Export audit trail dialog box
The audit trail exports the entire audit trail in formatted clear text with a checksum to ensure
data integrity.
Filename text box allows you to specify the filename for the exported data file. You may
specify a 20-character file name. The default file name is mmddyyyyhhmmss (month, day,
year, hour, minute, seconds) with the .txt extension.
Exporting record no: X thermometer displays the progress of the export.
Export icon starts the export. Before the export begins, the analyzer verifies that there is data
to export, that you have entered a valid file name, and that a USB flash drive is inserted into
the analyzer’s USB port. The file is written to a PAT700 directory in the root directory of the
flash drive. If the file already exists, you are asked if you want to overwrite it.
Return icon returns you to the audit trail dialog box.
Figure 5-23 Filter data dialog box