User manual

PAT700 Total Organic Carbon Analyzer – Setup Page 67
Anatel Operator Manual
To change the printer settings:
1) Turn the printer power switch OFF.
2) Restore power while pressing the ONLINE button. Release the button when a
printout of the current settings begins.
3) Push the ONLINE button again. “DIP SW1” is printed, prompting you for changes to
switches 1 through 8.
4) Enter the settings for DIP switch 1 by pressing ONLINE for ON (enabled) and FEED
for OFF (disabled) for switches 1 through 8 in succession. The ONLINE LED
illuminates to indicate an ON entry; the OFFLINE LED indicates an OFF entry.
Make sure to enter a setting for all eight switches. The setup mode cannot be
cancelled once initiated.
5) After all eight switches have been set, a printout prompts you whether to continue or
repeat the procedure for DIP switches 1 through 3 or write the current settings to
memory. After the settings have been saved, a printout is generated and the printer
returns to normal operation. Do not turn the printer off while it is saving settings.
Wait until “Dip SW setting complete!” is printed before shutting off power.
DIP Switch 3
Switch Setting Function
1 On 8 data bits
2 Off Parity setting = Yes
3 On Parity condition = Odd
4 Off Flow control (X on/X off)
5 Off 9600 baud rate
6 On 9600 baud rate
7 On 9600 baud rate
8 On 9600 baud rate
Table 5-2 Printer DIP switch settings