PAT700 Total Organic Carbon Analyzer – Setup Page 57
Anatel Operator Manual
Stabilization tab
Figure 5-5 TOC dialog box, stabilization tab
When the analyzer switches to perform an on-line TOC analysis, after running the analyzer in
continuous flow modes such as idle with conductivity, conductivity mode or standby for extend
periods of time, a thermal stabilization and cleaning of the analysis cell can reduce the
number of TOC analysis cycles necessary to produce stable results. An automatic
stabilization process, along with its activation and duration criteria, can be enabled in the
stabilization dialog box. The stabilization process involves trapping water in the cell and
turning on the UV lamp to thermally stabilize and clean the cell.
Stabilize cell prior to analyzing check box allows you to enable automatic stabilization of
the cell prior to performing a reportable TOC analysis.
Time between TOC analyses text box allows you to enter the minimum elapsed time
between TOC analyses required to activate the automatic stabilization process. You may
specify a time in hh:mm:ss format between 00:00:01 and 23:59:59. The default value is
02:00:00.
Stabilization time text box allows you to enter the amount of time to run the stabilization
process. You may specify a time in hh:mm:ss format between 00:00:01 and 23:59:59. The
default value is 00:10:00