Page 52 Run Modes – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
In manual self clean mode, the analyzer starts by setting the elapsed time to 0, opening the
cell valve to flush the cell, and turning the lamp on. If timed self cleaning was selected, the
dialog box displays the time remaining. Otherwise, the dialog box displays the elapsed time.
Self cleaning stops if you touch the stop icon (red “X”) or if the analyzer is switched to another
mode by schedule or by digital control.
Figure 4-9 Clean mode dialog box