Page 36 Installation – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Example:
Read command: “HR”
Read format:
“HH:mm:SS HH:mm:SS u u u u s u u u u u u u u u u u u
HH:mm:SS u u f u u u u u u f”
Response:
“00:01:00 00:00:00 50 0 1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00 25 00 1.000 000 020 005 095 0 0 1.000”
Write format:
“HR HH:mm:SS HH:mm:SS u u u u s u u u u u u u u u u u u
HH:mm:SS u u f u u u u u u f”
Write command:
“HR 00:01:00 00:00:00 50 0
1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00 25 00 1.000 000 020 005 095 0 0 1.000”
Example:
Read command: “HH”
Response:
“UVLamp1: 4 4300 1.000000 Good”
“UVLamp2: 0 4300 1.000000 Good”
Example:
Read command: “HC”
Response: “1.000”
Write format: “HC f”
Write command: “HC 0.95”
Example:
Read command: “HT”
Response: “1.000”
Write format: “HT f”
Write command: “HT 1.05”
“HH”
Get number of times UV lamp is turned on, hours left, voltage and status of both UV lamps
“HC”
Set/get user-calibrated
conductivity slope
Slope range should be between 0.9 and 1.1.
“HT”
Set/get user TOC slope Slope range should be between 0.85 and 1.15.