PAT700 Total Organic Carbon Analyzer – Installation Page 33
Anatel Operator Manual
Where:
Data logger commands
Example:
Read command: “RE”
Response:
Alarm time Alarm code
“21:38:15 Code #27”
“21:44:40 Code #36”
“21:49:05 Code #36”
“…”
“…”
Example:
Read command: “LU”
Response: “7%”
Mode State
1 = Online TOC 0 = Idle
2 = Manual TOC 1 = Sampling
3 = Digital TOC 2 = Oxidizing
7 = Online conductivity 4 = Flushing
11 = Offline
12 = StandBy
“SA”
Read data
Read-only command enables automatic result printouts from
the serial port, such as continuously streaming TOC data after
every analysis or conductivity data every 15 seconds. The read
format is same as “RD” command.
“RE”
Read alarm log Produces a list of alarms that occurred since power-up.
“ME”
Clear alarm log
Clears the alarm log list, also clears high TOC and conductivity
alarms.
“LE”
Erase alarm and data log
Erases contents of alarm and data logs. Issue this command to
reduce the size of the data log before taking a series of related
measurements.
“LU”
Data log percentage
Produces the percentage (0 to 100) of the data log that has
been used.