PAT700 Total Organic Carbon Analyzer – Installation Page 25
Anatel Operator Manual
2.6 Wiring connections for PAT700 with quick-connect fittings
Refer to this section if the PAT700 has five quick-connect fittings, as illustrated in Figure 2-13.
• One connector accommodates power supply wiring.
• The other four connectors accommodate 4-20 mA or discrete I/O wiring.
• The analyzer comes with factory-supplied mating connectors and a plug-in cord for
power supply wiring.
Figure 2-13 PAT700 power supply and I/O wiring terminals
Power supply wiring
WARNING
Potential property damage or electrical shock! Adhere to ground network
requirements for the facility.
WARNING
After power supply wiring has been installed, make sure the field wiring box cover is
in place and the screw is tightened with a screwdriver to ensure that the cover is
properly bonded to protective earth.
4-20 mA output wiring
connector
Digital input wiring
connector
Serial communications
wiring connectors
(see page 30)
Power supply wiring
connectors
Digital output wiring
connectors