Page 20 Installation – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Power supply wiring
WARNING
Potential property damage or electrical shock! Adhere to ground network
requirements for the facility.
Figure 2-10 illustrates power supply wiring terminals.
Table 2-2 lists specifications for the power supply.
Select a wire style and gauge that meets local electrical codes. The power supply wiring
connector accepts 10 to 14 AWG wiring. Wire type can be stranded or solid. Strip 3/8”
(9.5 mm) of insulation from the ends of the wires and insert the stripped wire ends into the
appropriate connector terminals. Secure the wire with the integral screw and tug firmly on
each wire to make sure it is properly secured.
WARNING
Potential electrical shock! After power supply wiring has been installed, make sure the
field wiring box cover is in place and the screw is tightened with a screwdriver to
ensure that the cover is properly bonded to protective earth.
Figure 2-10 PAT700 power supply wiring terminals
Table 2-2 Power supply specifications
Description Specification
Voltage input 100 to 230 VAC universal
Frequency 50 to 60 Hz
Power 65 W maximum, fused at 1.25 A
Line
Neutral
Ground