PAT700 Total Organic Carbon Analyzer – Specifications Page 177
Anatel Operator Manual
Input/output specifications
Analog output specifications
Analog outputs Three, opto-isolated
Variables TOC, conductivity, and temperature
Range Configurable over 4-20 mA
Alarm level Configurable: 2, 22, or last value
Discrete input specifications
Discrete inputs Two, opto-isolated
Input one TOC analysis start
Input two TOC/Flow with conductivity mode change
Discrete output specifications
Discrete outputs Four, opto-isolated
Output one TOC alarm
Output two Uncompensated conductivity alarm
Output three Analyzer error
Output four TOC analysis start indication
Serial output specifications
Serial RS-232C output ASCII format, 8 data bits, no parity, 1 stop bit, 9600 baud
Serial printer output RS-232C interface, 8 data bits, no parity, 1 stop bit, 9600 baud
Compatibility 40-column serial printer
Ethernet output specifications
Protocol Modbus TCP/IP
Interface speed 10/100 Mbits/sec.
USB output specifications
USB output One, output to USB flash drive