Page 176 Specifications – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Physical specifications
UV lamps Two, with UV Detect technology
Interface/display Color touch screen
Maximum altitude 4,000 m (13,125 ft)
User I/O wiring Three 3/4-inch conduit openings or quick disconnect fittings
Standards system Onboard, automated standards introduction system (OASIS)
Dimensions 59.7 cm wide x 22.9 cm deep x 25.4 cm high
(23.5” wide x 9” deep x 10” high)
Weight 13.6 kg (30 lbs)
Sample inlet flow rate range 60 mL/min to 300 mL/min
Sample inlet pressure range 10 to 100 psi
Electronics enclosure rating • IP56 conduit version
• IP46 quick connect version
Humidity rating 5 to 95%
CE compliance EN61010-1 and EN 61326
Safety rating ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Protection and overvoltage Category II
Pollution degree 2, IEC 61010-1
Power input 100 to 230 ±10% VAC, 65 W, 50/60 Hz