PAT700 Total Organic Carbon Analyzer – Installation Page 9
Anatel Operator Manual
Chapter 2 Installation
DANGER
Tasks involving installation of this equipment have fire, electrical, and pressure related
hazards associated with them. These tasks must only be attempted by individuals
trained and knowledgeable in the particular task and the associated hazards.
2.1 Installation requirements
The quick-connect version of the PAT700 has an IP46 enclosure surrounding the electronics
portion of the analyzer. The conduit version of the PAT700 has a waterproof IP56 enclosure,
that can withstand water jets, surrounding the electronics portion of the analyzer. The
hydraulics portion of the analyzer is protected by a splash resistant enclosure, but does not
carry an IP rating.
Install the analyzer in a dry, dust-free environment.
Install the analyzer in a properly ventilated location where the ambient air temperature
is 10 to 40 °C (50 to 104 °F).
The analyzer has 3/4-inch NPT conduit fittings or quick-connect fittings for wiring.
See Figure 2-1, page 11 for mounting dimensions. The analyzer weighs 30 lb (13.6 kg).
The analyzer ship kit includes the following parts:
This instruction manual
5-foot (1.5 meter) PFA inlet tubing
10-foot (3 meter) polypropylene outlet tubing
5/16-inch T-handle Allen wrench
•Stylus
Instrument screwdriver
USB memory stick
Graduated cylinder
For units with quick connect I/O fittings the following items are also included:
Quick connect power cord
Male 8-pin quick connect cord fitting
Three male 5-pin quick connect cord fittings