Page 156 Modbus Protocol – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Address 136-139 (Cell calibration offset) (Format: Double Float)
Get the instrument's Cell calibration offset. The Cell calibration gain consists of 4 words and
read as 64bit double float data.
Address 140-143 (TOC slope) (Format: Double Float)
Get the instrument's TOC slope. The TOC slope consists of 4 words and read as 64bit double
float data.
Address 144-147 (Conductivity slope) (Format: Double Float)
Get the instrument's Conductivity slope. The Conductivity slope consists of 4 words and read
as 64bit double float data.
Address 148-151 (Thermistor coefficient A) (Format: Double Float)
Get the instrument's Thermistor coefficient A. The Thermistor coefficient A consists of 4 words
and read as 64bit double float data.
Address 152-155(Thermistor coefficient B) (Format: Double Float)
Get / Set the instrument's Thermistor coefficient B. The Thermistor coefficient B consists of 4
words and read as 64bit double float data.
Address 156-159 (Thermistor coefficient C) (Format: Double Float)
Get the instrument's Thermistor coefficient C. The Thermistor coefficient C consists of 4
words and read as 64bit double float data.
Address 160 (TOC Print) (Format: 16 Bit Word)
Get / Set the instrument's TOC Print. The TOC Print consists of 1 word which refers to 0 =
Disable, 1 = Continuous and 2 = % Change.
Address 161 (TOC Print Percentage change) (Format: 16 Bit Word)
Get / Set the instrument's TOC Print Percentage change. The TOC Print Percentage change
consists of 1 word.
Address 162 (Conductivity Print) (Format: 16 Bit Word)
Get / Set the instrument's Conductivity Print. The Conductivity Print consists of 1 word which
refers to 0 = Disable, 1 = Interval and 2 = % Change.
Address 163-164 (Conductivity Print time) (Format: Time 2)
Get / Set the instrument's Conductivity Print time. The Conductivity Print time consists of 2
words.
Address 165 (Conductivity Print Percentage change) (Format: 16 Bit Word)
Get / Set the instrument's Conductivity Print Percentage change. The Conductivity Print
Percentage change consists of 1 word.