Page 154 Modbus Protocol – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Address 102-103 (Subnet Mask) (Format: IP Address)
Get the instrument's Subnet Mask (Manual / Auto DHCP). The Subnet Mask consists of 4
bytes of binary characters (0x00-0xFF). Each 16-bit register contains two 8-bit characters.
Address 104-105 (Gateway) (Format: IP Address)
Get the instrument's Gateway (Manual / Auto DHCP). The Gateway consists of 4 bytes of
binary characters (0x00-0xFF). Each 16-bit register contains two 8-bit characters.
Address 106-107 (DNS Server) (Format: IP Address)
Get the instrument's DNS Server (Manual / Auto DHCP). The DNS Server consists of 4 bytes
of binary characters (0x00-0xFF). Each 16-bit register contains two 8-bit characters.
Address 108-109 (WINS Server) (Format: IP Address)
Get the instrument's WINS Server (Manual / Auto DHCP). The WINS Server consists of 4
bytes of binary characters (0x00-0xFF). Each 16-bit register contains two 8-bit characters.
Address 110 (DHCP Flag) (Format: 16 Bit Word)
Get the value that states the instrument's TCP/IP connection type i.e. Either Manual or Auto
DHCP. The DHCP Flag consists of 1 word which refers to 0 = Manual and 1 = Auto DHCP.
Address 111 (Language Identification code) (Format: 16 Bit Word)
Get / Set the value that states the instrument's language type. The Language identification
code consists of 1 word which refers to 0 = English, (Future implementation: 1 = French, 2 =
German, 3 = Italian, 4 = Spanish, 5 = Chinese Simplified and 6 = Japanese.)
Address 112 (Mode of Operation) (Format: 16 Bit Word)
Get / Set the instrument's Mode of Operation. The Mode of Operation consists of 1 word
which refers to 0 = Offline, 1 = Online TOC, 2 = Online Conductivity, 3 = Digital TOC, 4 =
Manual TOC, 5 = Manual Bottle, 6 = Manual Clean, 7 = Manual Diagnostics and 8 = Standby.
NOTE
Modes 5 = Manual Bottle and 7 = Manual Diagnostics can only be read and not set.
Address 113 (Current State) (Format: 16 Bit Word)
Get the value that refers to the instrument's current operational state. The Current State
consists of 1 word which refers to 0 = Idle, 1 = Flushing, 2 = Analyzing, 3 = Clean, 4 =
Standby, 5 = Offline, 6 = Error, 7 = Bottle mode and 8 = Diagnostic mode.
Address 114-115 (Cycle Time) (Format: Time 1)
Get / Set the instrument's Cycle Time. The Cycle Time consists of 2 words.
Address 116-117 (Flush Time) (Format: Time 1)
Get / Set the instrument's Flush Time. The Flush Time consists of 2 words.