User manual

PAT700 Total Organic Carbon Analyzer – Data Review Page 137
Anatel Operator Manual
Chapter 9 Data Review
9.1 Accessing the data review dialog box
1) At the home screen, touch the << icon to open the sliding toolbar.
2) Touch the data review icon.
Access to the data review dialog box may be delayed while available data is transferred to the
data log.
9.2 Data review dialog box
The data review dialog box is populated only when data logging is selected in the security
settings dialog box. The data log will hold up to 5,000 records. The data log contains analysis
results, alarms, bottle test results, and audit trail entries. Since the audit trail is stored in the
data log each audit trail entry uses one of the 5,000 data log entries. The data log operates in
a first in first out approach. Once the data log contains 5,000 entries the oldest entry is
replaced with the newest entry. See Security, page 69.
Figure 9-1 Data review dialog box
Data review window lists all the data in the data log. The data appear as they would on the
printer. The data are filtered based on the settings in the filter dialog box. Data are displayed
in chronological order from earliest to most recent. The list scrolls to display the earliest data.
Print icon sends the displayed data to the printer.
Export icon opens the export settings dialog box. See Export data, page 138.
Return icon returns you to the home screen.
Filter data icon opens the filter data dialog box. See Filter data, page 139.