User manual
Page 134 Diagnostics – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Figure 8-14 Fan test dialog box
RFID test
The RFID test icon allows you to test the RFID reader system.
To conduct this test you will need to install four standards bottles equipped with RFID tags.
These bottle can be new or used standards bottles.
Open the OASIS system door and insert up to four standards bottles with RFID tags in each
bottle location. Touching the start button will initiate a test of the RFID system. The system will
automatically sense those locations where RFID tags are present. To verify failure of the
system the unlock button can be used to open the OASIS door during the test. With the door
open, and the bottles no longer positioned properly, the system will return a failed test.
Figure 8-15 RFID test dialog box
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to open the
door.
Return icon returns you to the diagnostics dialog box.
Run icon initiates the RFID test.