Page 132 Diagnostics – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
6) If a test is chosen immediately, the analyzer asks if you wish to install a bottle. If yes
is selected, the load bottles dialog box illustrates the appropriate port in which to
install the bottle. The enter bottle information screen requires you to enter the
bottle's volume. The next button then starts the test.
Pump test
The pump test icon allows you to test pumps. See Figure 8-12.
Before the pump test, you must place a 50 mL graduated cylinder at the analyzer's water
outlet. No tubing should be used. A bottle must be installed in position 2 to enable the pump
to be primed. When you touch the run icon, the analyzer pumps water for 30 seconds, then
prompts you to input the volume pumped, as measured in the graduated cylinder. After you’ve
input the volume, touch the return icon to accept the results or the cancel icon to reject them.
Figure 8-12 Pump test dialog box
Calibrate pump icon opens the calibrate pump dialog box. See Figure 8-13.
Start/stop pump icon starts and stops the pump. Touching the start pump icon opens the
pump valve and the grab sample bottle valve, then starts the pump. The pump runs
indefinitely until you press the stop pump icon. Both valves then close, and the pump stops.
Prime pump icon opens the pump valve and the grab sample valve, then starts the pump.
The pump runs long enough to prime, then it stops. Both valves then close.
Load bottles icon opens the load bottles dialog box. See Figure 8-10, page 131.
Unload bottles icon takes you to the unload/replace bottles dialog box. See Figure 8-11,
page 131.
Return icon returns you to the diagnostics dialog box.