Page 120 Alarms – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
SC104 NotEnoughFluidinBottle1 • Not enough fluid in bottle 1 to complete the operation.
• Check bottle 1 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
SC105 NotEnoughFluidinBottle2 • Not enough fluid in bottle 2 to complete the operation.
• Check bottle 2 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
SC106 NotEnoughFluidinBottle3 • Not enough fluid in bottle 3 to complete the operation.
• Check bottle 3 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
SC107 NotEnoughFluidinBottle4 • Not enough fluid in bottle 4 to complete the operation.
• Check bottle 4 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
SC108 UnableToReadSettings Some or all of the settings could not be read from the settings file.
SC109 UnableToWritetoSettings Analyzer can not write to the settings file.
SC118 UnableToChangeLanguageSettings Analyzer can not change the language settings.
SC119 UnableToReadLanguageSettings Analyzer can not read the language settings
SC120 UnableToChangeEthernetSettings Analyzer can not change the Ethernet settings.
SC121 UnableToReadEthernetSettings Analyzer can not read the Ethernet settings.
SC122 UnableToPrintSystemSettings Analyzer can not print the system settings.
SC125 UnableToStartLampTest Analyzer can not start the lamp test.
SC126 UnableToStopLampTest Analyzer can not stop the lamp test.
SC127 UvLampLifeExceeded • Lamp hours of usage have been exceeded or the UV Detect™
output has dropped below the acceptable limit.
• Lamp has been set to a status of marginal.
• Replace the lamp.
SC129 UnableToReadInformationFromBottle Analyzer can not read the information from the bottle.
SC130 UnableToWriteInformationToBottle Analyzer can not write the information to the bottle.
SC132 UnableToStartPumpTest Analyzer can not start the pump test.
SC133 UnableToStopPumpTest Analyzer can not stop the pump test.
SC134 UnableToCalibratePump Analyzer can not calibrate the pump.
SC135 UnableToExportDataLog Analyzer can not export the data log.
SC136 UnableToExportAuditTrail Analyzer can not export the audit trail.
SC137 UnableToAddUser Analyzer can not add a user.
SC138 UnableToEditUser Analyzer can not edit a user.
SC139 UnableToDeleteUser Analyzer can not delete a user.
Table 7-3 Warning alarm messages
Code Message Alarm condition