PAT700 Total Organic Carbon Analyzer – Alarms Page 119
Anatel Operator Manual
Critical error alarms
A critical error occurs any time the analyzer cannot function due to an error in the system. The
error appears as a full-screen, flashing message. All operation ceases until the error condition
has been corrected. The current analysis is aborted. Touching the screen provides access to
the alarm acknowledge dialog box. See Table 7-2.
Warning alarms
A warning occurs any time one of the subsystems is not operating at a normal level. When a
warning condition is present, a warning indicator appears on the display. Operation may
continue with a warning condition present. When the warning condition has been corrected,
the indicator disappears. See Table 7-3.
Table 7-2 Critical error alarm messages
Code Message Alarm condition
SC110 UnableToStartIoModule Analyzer can not start the IO module.
SC111 UnableToStartTocModule Analyzer can not start the IO module.
SC114 UnableToStartEthernetModule Analyzer can not start the Ethernet module.
SC152 LampUnableToSwitch Analyzer can not switch from one lamp to the other.
SC173 RangeTableFailedToLoad RangeTable.txt file could not be loaded.
SC174 InstrumentTemperature RangeTable.txt file could not be loaded.
SC175 BothLampsBad Both UV lamps have failed.
Replace both lamps as soon as possible.
SC176 NoCondTempReadings Analyzer can not measure conductivity and temperature.
Restart the analyzer.
If restarting does not resolve the error, phone customer
service at 1–800–866–8854.
SC179 HardwareInitializationFailure An error occurred during hardware initialization.
Table 7-3 Warning alarm messages
Code Message Alarm condition
SC100 IncorrectBottleLoadedinPosition1 Incorrect bottle detected in position 1.
Analyzer can not perform bottle test operation.
SC101 IncorrectBottleLoadedinPosition2 Incorrect bottle detected in position 2.
Analyzer can not perform bottle test operation.
SC102 IncorrectBottleLoadedinPosition3 Incorrect bottle detected in position 3.
Analyzer can not perform the bottle test operation.
SC103 IncorrectBottleLoadedinPosition4 Incorrect bottle detected in position 4.
Analyzer can not perform the bottle test operation.