Page 118 Alarms – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Error alarms
Errors occur any time one of the non-critical subsystems fail or a failure occurs from which the
analyzer attempts to recover. The animation area flashes red to indicate an error condition is
present. When an error condition is present, the analyzer attempts to recover and continue
operating. See Table 7-1.
Table 7-1 Error alarm messages
Code Message Alarm condition
SC027 FalseConductivity Measured conductivity is less than 0.
SC029 BadTocConvert Temperature/Conductivity is out of bounds for TOC conversion.
SC035 LowOxidationRate TOC is too low (out of range).
SC036 ChemicalInterference Negative TOC (out of range)
Potential presence of organic acids
SC037 LowTemperature Temperature is out of range low.
SC038 HighTemperature Temperature is out of range high.
SC039 LowResistance Resistivity is out of range low.
SC040 HighResistance Resistivity is out of range high.
SC112 UnableToStartPrinterModule Analyzer can not start the printer module.
SC113 UnableToStartRs232Module Analyzer can not start the RS-232 module.
SC116 UnableToReadDataLog Analyzer can not read the data log.
SC117 UnableToWriteToDataLog Analyzer can not write to the data log.
SC128 UvLampMonitorFailed UV lamp monitor has failed.
SC151 LampExtinct Analyzer detects that the current lamp has failed.
Replace the UV lamp.
SC163 UnableToStopCurrentOperation Analyzer can not stop its current operation.
SC164 UnknownError Analyzer has experienced an unknown error.
SC167 WatchDogIoCard I/O Card has stopped responding.
SC168 WatchDogTocAlgoithm TOC algorithm has stopped responding.
SC169 WatchDogPrinter Printer has stopped responding.
SC170 WatchDogRs232 RS-232 module has stopped responding.
SC177 Exception An exception was thrown.
SC188 TOCOverLimit TOC Over limit has occurred.