User manual

PAT700 Total Organic Carbon Analyzer – Alarms Page 115
Anatel Operator Manual
Chapter 7 Alarms
7.1 Alarm indication
The analyzer has two basic operational modes: on-line and bottle. On-line TOC and
conductivity are the two primary on-line modes (although you can also select standby, off-line,
manual, or self clean). In on-line TOC and conductivity modes, the home screen provides an
on-screen animation indicating the general state of water flow through the analyzer. In on-line
TOC, conductivity, or manual TOC modes, the animation area also serves as an alarm
condition indicator. If an alarm occurs in an on-line mode, the following sequence of events
occurs:
When a TOC or conductivity alarm condition occurs, the home screen animation area
flashes red, the displayed value of the variable causing the alarm also turns red, the
PAT700 makes a sound (if a sound is programmed), the alarm icon is enabled, and the
alarm condition is written to the data log and audit trail. See Figure 7-1.
Figure 7-1 Online TOC mode, current tab, in alarm condition
The animation area continues to flash red until a user acknowledges the alarm. See
Alarm acknowledgement, page 116.
If the alarm has been acknowledged and the user returns to the home screen, the
animation area remains solid red until the analyzer completes a full cycle without any
errors.
Once a full cycle has occurred without any errors, the normal animation reappears.
If on the next cycle the same or a new error occurs, the animation area again flashes
red and must be acknowledged.
As long as alarms exist in the alarm log, the alarm log is accessible.